THERMAL SHOCK CHAMBER – VHAGAR SERIES
THERMAL SHOCK TEST CHAMBER – VHAGAR SERIES
The VHAGAR series is the DOT platform dedicated to rapid thermal shock testing, designed to subject materials, components, and devices to extreme and repeated temperature variations.
VHAGAR chambers allow severe thermal stresses to be simulated in accordance with the main international standards:
- IEC 60068-2-14
- MIL-STD-810 (Method 503)
- MIL-STD-883
They are intended for test laboratories, R&D departments, and validation in the automotive, aerospace, defense, electronics, and telecommunications sectors.
VHAGAR chambers are available in:
-
Dynamic configuration (2-zone system)
With a motorized basket transferring the specimen between hot and cold chambers. -
Static configuration (3-zone system)
With a fixed test area and thermal shock generated by controlled switching of hot and cold airflow (required when the DUT must remain electrically or mechanically connected).
The system is equipped with a proprietary DOT 10” Touch Panel with ClimaLogic® software.
OPERATING RANGE
Temperature Range
-
−75 °C to +200 °C
-
Available versions:
-
A: −45 °C
-
B: −55 °C
-
C: −75 °C
-
Dynamic Performance
-
Specimen transfer time (2-chamber dynamic version): ≤ 10 s
-
Stabilization time in test zone (3-chamber static version): ≤ 5 min
-
Heating time hot zone (RT → +200 °C): approx. 35 min
-
Cooling time cold zone (RT → −70 °C): approx. 55 min
Accuracy and Stability
-
Control accuracy: ±0.5 °C
-
Temperature uniformity: ±2.5 °C
STANDARD CONFIGURATION
-
10" DOT touchscreen with adjustable arm support
-
Industrial PLC for system management and safety interlocks
-
8 auxiliary digital outputs
-
2 adjustable shelves (≥ 30 kg load per shelf)
-
Inspection window on hot chamber (optional on cold chamber) with LED lighting
-
Ø100 mm cable feedthroughs on hot and cold zones
-
Top cable guide feedthrough
-
Cycle start selectable from hot or cold zone
-
Independent hot-only or cold-only operation mode
-
Independent safety thermostat
-
Ethernet port
-
Castors and leveling feet
STRUCTURE AND MATERIALS
-
External structure: SUS 304 stainless steel with brushed finish
-
Internal test space: SUS 304 stainless steel with mirror finish
-
Insulation: high-density rigid polyurethane foam + rock wool
-
Double-layer door sealing system
-
High-quality industrial surface finishing
REFRIGERATION SYSTEM
-
Two-stage cascade refrigeration system
-
Copeland or Bitzer compressors
-
Water-cooled condenser (standard configuration)
-
Air-cooled condenser (available upon technical evaluation)
CONTROL SYSTEM
-
DOT 10" Touch Panel with ClimaLogic® software
-
PID control with SSR management and dedicated industrial microprocessor
-
Up to 500 programmable cycles (20 segments each)
-
Cycle concatenation up to 160 segments
-
Cycle repetition up to 9,999 times
-
Setpoint modification during operation
-
Continuous data logging
-
.log ASCII file export (Excel compatible)
-
Real-time graphical monitoring
-
Remote access via LAN (Dydrus software)
AVAILABLE MODELS
VHAGAR – 2-ZONE THERMAL SHOCK
Available models:
-
Vhagar 42
-
Vhagar 80
-
Vhagar 150
-
Vhagar 200
-
Vhagar 300
Usable capacity: 42 L to 300 L
Power supply: 400 V ±10% / 50 Hz / 3 + N + G
Weight: 500 kg to 1400 kg (depending on model)
VHAGAR – 3-ZONE THERMAL SHOCK
Available models:
-
Vhagar 42
-
Vhagar 80
-
Vhagar 100
-
Vhagar 150
-
Vhagar 300
-
Vhagar 450
Usable capacity: 42 L to 450 L
Power supply: 400 V ±10% / 50 Hz / 3 + N + G
Weight: 700 kg to 1900 kg (depending on model)
AVAILABLE OPTIONS
-
Air-cooled condenser
-
Electrical disconnection interface
-
GN2 or dry air purge system
-
LN2 or CO₂ cooling booster
-
Noise reduction kit (−3 dB(A))
-
Variable airflow speed in cold chamber (ESS mode)
-
DUT temperature monitoring sensor
-
Top cable guide system with feedthrough
-
Custom configurations on request
TECHNICAL POSITIONING
The VHAGAR series represents the DOT solution for high-performance thermal shock testing, offering both dynamic and static configurations, advanced ClimaLogic® control and industrial-grade architecture designed for severe, high-intensity and fully repeatable thermal shock cycles over time.
Request a custom configuration tailored to your industry!
AT YOUR FULL DISPOSITION
Request information or a FREE quote