Monoblock Climatic Chambers for Stability Studies

Monoblock Climatic Chambers

Solutions for Pharmaceutical and Cosmetic Stability Studies

The monoblock climatic chambers within the Il Disgelo Pharma & Cosmetics line are systems designed for pharmaceutical and cosmetic stability studies, providing precise and continuous control of temperature and relative humidity.

The equipment is designed and manufactured by Caelis in France and distributed in Italy by Il Disgelo, which provides installation, qualification, maintenance and technical support throughout the national territory.

The monoblock configuration integrates the main climatic treatment and control systems into a single structure, allowing simple and rapid installation in stability laboratories and ensuring continuous 24/7 operation during long-term studies.

Air distribution is designed to maintain uniform and repeatable climatic conditions even under loaded conditions, an essential requirement for stability studies, qualification activities and controlled storage.

The range is available in different usable volumes, from 560 to 1,998 L, and supports the main climatic conditions commonly used in stability studies according to ICH Q1A guidelines.

The chambers are designed to provide:

  • precise temperature and humidity control;
  • stable and repeatable climatic conditions;
  • continuous 24/7 operation;
  • uniform air distribution even under load;
  • long-term stability testing;
  • continuous monitoring of climatic parameters;
  • recording and traceability of test data;
  • integrated alarm management;
  • programmable climatic cycles;
  • configurations dedicated to pharmaceutical and cosmetic laboratory requirements.
Technical specifications.

‎ ‎ ‎ ‎

  • type: monoblock climatic chambers for stability studies;
  • application: pharmaceutical and cosmetic sectors;
  • operation: designed for continuous 24/7 use;
  • available usable volumes:560 L, 1,300 L and 1,998 L;
  • temperature range:+5 °C to +50 °C;
  • temperature stability:< 0.5 °C;
  • temperature uniformity:±2 °C under load;
  • relative humidity range:20% to 85% RH;
  • humidity stability:< 1% RH;
  • humidity uniformity:±5% RH under load;
  • main ICH conditions: 25 °C / 60% RH, 30 °C / 65% RH, 30 °C / 75% RH, 40 °C / 75% RH;
  • internal chamber: stainless steel;
  • control:7” touch screen with multilingual interface;
  • technical feedthroughs:Ø16 mm and Ø80 mm;
  • anti-condensation system: integrated;
  • mobility: casters with brakes;
  • water tank:20 L;
  • monitoring: continuous temperature and humidity monitoring;
  • data logging: recording and traceability of climatic parameters;
  • alarm management: integrated;
  • programming: programmable climatic cycles.

Available Models

MODELUSABLE VOLUMEHEIGHTWIDTHDEPTHNUMBER OF SHELVES
CP-600-C560 L1985 mm750 mm820 mm7
CP-1300-C1300 L1985 mm1500 mm820 mm14
CP-2000-C1998 L1985 mm2250 mm820 mm21

‎ ‎ ‎ ‎

Configurable Data

The values shown represent technical configurations of the monoblock range. Usable volume, number and arrangement of shelves, internal configuration, technical feedthroughs and optional equipment may vary depending on the selected model and the specific requirements of the laboratory.

The final configuration is defined according to product type, quantity of samples, required climatic conditions, study duration, monitoring method and qualification requirements.

Options

‎ ‎ ‎

  • Internal LED lighting
  • Additional shelves or dedicated internal configurations
  • Additional technical feedthroughs
  • Supplementary monitoring systems
  • Dedicated solutions for data recording and traceability
  • Specific configurations for different stability conditions
  • Installation and commissioning
  • IQ – Installation Qualification
  • OQ – Operational Qualification
  • PQ – Performance Qualification
  • Preventive maintenance
  • Specialized technical service
  • Periodic performance verification

Note: available options, qualification services and configurations may vary depending on the model, application and laboratory requirements.

Request information

At your complete disposal for any information.

Please enter in the fields below, all the requested data. We will respond to your request as quickly as possible.